TITLE 32: ENERGY
CHAPTER II: ILLINOIS EMERGENCY MANAGEMENT AGENCY
SUBCHAPTER b: RADIATION PROTECTION
PART 335 MEDICAL USE OF RADIOACTIVE MATERIAL
SECTION 335.8230 THERAPY-RELATED COMPUTER SYSTEMS FOR REMOTE AFTERLOADER UNITS, INTRAVASCULAR BRACHYTHERAPY UNITS, TELETHERAPY UNITS AND GAMMA STEREOTACTIC UNITS


 

Section 335.8230  Therapy-related Computer Systems for Remote Afterloader Units, Intravascular Brachytherapy Units, Teletherapy Units and Gamma Stereotactic Units

 

The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies.  At a minimum, the acceptance testing must include, as applicable, verification of:

 

a)         The source-specific input parameters required by the dose calculation algorithm;

 

b)         The accuracy of dose, dwell time and treatment time calculations at representative points;

 

c)         The accuracy of isodose plots and graphic displays;

 

d)         The accuracy of the software used to determine sealed source positions from radiographic images; and

 

e)         The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.

 

(Source:  Added at 30 Ill. Reg. 9029, effective April 28, 2006)